Content
The content of medical knowledge in this section of the site of the Lactology Foundation is intended for the practical needs of doctors, pharmacists and students in these specialties. It is more than reasonable to consult other authoritative medical sources before using our medical knowledge.
- Lung in Health and Disease
- General Approach to Patients
With Respiratory Disorders - Evaluating Lung Structure and Function
- Interstitial Lung Diseases
- Pulmonary Vascular Diseases
- Disorders of the Pleura,
Mediastinum, and Chest Wall - Respiratory Failure
- Lung Transplantation
- Perioperative Pulmonary Management
- COVID-19 Pulmonary Management
- Congenital Lung Malformations
- Sleep-Related Disorders
Polysomnography
Nocturnal, laboratory-based polysomnography (PSG), also known as a sleep study, is the most commonly used test in the diagnosis of obstructive sleep apnea syndrome (OSAS).
Parameters Monitored
Assessment of sleep
stages requires 3 studies:
electroencephalography (EEG), electrooculography
(EOG), and surface electromyography (EMG).
One EEG channel (central channel with an ear
reference provides the best amplitude) is used
to monitor sleep stage. However, most
laboratories use 2 central channels and 2
occipital channels, with ear references as an
adjunct to help identify sleep latency and
arousals. A 10- to 20-electrode placement system
is used to determine the location of these
channels. Additional EEG channels can be used,
particularly in patients with epilepsy (ie, a
full 10-20 montage).
Two EOG channels are
used to monitor both horizontal and vertical eye
movements. Electrodes are placed at the right
and left outer canthi, one above and one below
the horizontal eye axis. The electrodes pick up
the inherent voltage within the eye; the cornea
has a positive charge and the retina has a
negative charge. Evaluation of the eye movements
is necessary for 2 reasons. First is for
documentation of the onset of rapid eye movement
(REM) sleep, and second is to note the presence
of slow-rolling eye movements that usually
accompany the onset of sleep.
One EMG channel
(usually chin or mentalis and/or submentalis) is
used to record atonia during REM sleep or lack
of atonia in patients with REM-related
parasomnias. To assess bruxism, the EMG
electrodes can be placed over the masseter. The
EMG recording from other muscle groups is
assessed for other sleep disorders. For example,
the anterior tibialis EMG is helpful for
assessing periodic limb movements during sleep
and the intercostal EMG is used as adjunctive
help for determining effort during respiratory
events.
Two channels are used for monitoring
airflow. One thermistor channel (oral and/or
nasal) is used to evaluate the presence or
absence of airflow. Any change in temperature as
a patient inhales and exhales leads to a normal
signal, so this channel is insensitive to
partial flow obstruction. Thermistor is the
recommended channel for evaluation of apneas.
Nasal pressure transducer channel is a more
sensitive measure of airflow restriction. Normal
breathing has a rounded pattern, while
resistance to airflow leads to a squaring off of
the flow signal. Pressure transducer is the
recommended channel for evaluating hypopneas. It
is also used for airflow resistance in upper
airway resistance syndrome. Other parameters
that can be monitored in a sleep study include
the following:
• Electrocardiography
•
Pulse oximetry
• Respiratory effort (thoracic
and abdominal)
• End tidal or transcutaneous
C02
• Sound recordings to measure snoring
• Surface EMG monitoring of limb muscles (to
detect limb movements, periodic or other)
•
Continuous video monitoring
Optional
parameters that can be monitored in a sleep
study include the following:
• Core body
temperature
• Incident light intensity
•
Penile tumescence
• Pressure and pH at
various esophageal levels
Staging of Sleep
Standardized criteria for
the staging of sleep were published first in
1968 by Rechtschaffen and Kales. A revised
version was published in 2007 by the American
Academy of Sleep Medicine. The chief revision
was the consolidation of stages 3 and 4 into a
single stage N3 (slow wave sleep). Previous
stages 1 and 2 were renamed N1 and N2. Both
systems are reflected below, with the
expectation that the discontinued stages will be
omitted from future revisions of this article.
EEG background
Alpha EEG
• Frequency of
8-13 Hz
• Produced in the occipital region
• Crescendo-decrescendo appearance Theta EEG
• Frequency of 3-7 Hz
• Produced in the
central vertex region
• No amplitude criteria
• Most common sleep frequency Delta EEG
• EEG
frequency of 0.5-2 Hz
• Seen predominantly in
the frontal region
• Amplitude of greater
than 75 microvolts
Theta EEG
• Frequency of 3-7 Hz
•
Produced in the central vertex region
• No
amplitude criteria
• Most common sleep
frequency Delta EEG
• EEG frequency of 0.5-2
Hz
• Seen predominantly in the frontal region
• Amplitude of greater than 75 microvolts
Delta EEG
• EEG frequency of 0.5-2 Hz
•
Seen predominantly in the frontal region
•
Amplitude of greater than 75 microvolts
Sleep spindle
See the list below:
•
Frequency of 12-14 Hz
• Produced in
central-vertex region
• Greater than 0.5-3
seconds in duration
• 0.5-second spindles
with 6-7 cycles
• Indicative of stage 2 sleep
K complexes
See the list below:
•
Sharp, slow waves with a negative, then
positive, deflection
• No amplitude criteria
• Duration must be at least 0.5 seconds
•
Predominantly produced in central-vertex region
• Indicative of stage 2 sleep
• May occur
with or without stimuli
Wake stage
See the list below:
•
Greater than 50% of each epoch contains alpha
activity
• Eye blinks at a frequency of 0.5-2
Hz
• Reading eye movements
• Irregular
conjugate rapid eye movements associated with
normal or high chin tone.
Stage N1 (formerly stage 1)
See the list
below:
• Greater than 50% of the epoch
contains theta activity (4-7 Hz) with slowing of
the background rhythms greater than or equal to
1 Hz from those of stage wake
• Vertex sharp
waves
• Slow-rolling eye movements in EOG
channels
• Relatively high submental EMG tone
Stage N2 (formerly stage 2)
See the list
below:
• Theta activity (4-7 Hz)
•
K-complexes and sleep spindles occur
episodically
• High tonic submental EMG
Stage N3
See the list below:
• Greater
than 20% of each epoch must contain delta
activity
• Amplitude of 75 microvolts or
greater
• Submental muscle tone may be
slightly reduced
Discontinued former stage 3
See the list
below:
• Between 20-50% of each epoch must
contain delta activity
• Amplitude of 75
microvolts or greater
• Submental muscle tone
may be slightly reduced
Discontinued former stage 4
See the list
below:
• Greater than 50% of the epoch has
scorable delta activity
• Amplitude of 75
microvolts or greater
• Submental EMG
activity slightly reduced from that of light
sleep
REM sleep
See the list below:
• Rapid
eye movements
• Low amplitude, mixed
frequency EEG (similar to awake pattern)
•
Atonia or the lowest tonic submental EMG
•
May see saw-tooth waves
Procedures
In 1992, the Office of
Technology Assessment of the Agency of Health
Care Policy and Research recommended, in an
evidence-based assessment, declared two tests as
having been studied sufficiently. Both tests are
performed in a sleep laboratory.
The first is
overnight polysomnography (PSG) or sleep study,
which is an overnight recording of the patient's
sleep. Typically for a baseline study, the
patient is observed sleeping naturally without
any treatment, but if a significant amount of
sleep-disordered breathing (AHI> 20-30 events
per hour) is seen in the first hours of the
study, a split-night study is performed during
which positive airway pressure (PAP) is started.
Titration studies may also be done with
initiation of PAP from the start of the study to
determine optimal settings.
The second is
multiple sleep latency testing (MSLT), which
records multiple naps throughout a day.
Maintenance of wakefulness testing (MWT) can
also be performed, which determines how long
wakefulness can be maintained.
Standard sleep
studies usually use the overnight PSG (may be
performed over several nights). If daytime
sleepiness is an issue and cannot be fully
explained by the overnight study results, an
MSLT should be performed the next day.
Limitations usually stem from the fact that
recording conditions may not reflect what
happens during a regular night in the patient's
home.
Although diagnosing a sleep problem on
the basis of a recording over a single night is
common practice, some authorities caution that
more than one night of recording may be
necessary so the patient can become comfortable
with unfamiliar surroundings and sleep more
naturally. This effect is greatest on the first
night in the sleep laboratory (ie, first-night
effect).
Sporadic events may be missed with a
single-night PSG. External factors that disturb
the subject's sleep may be present in the home
but absent from the controlled environment of
the sleep laboratory.
Patient preparation is important so that the
patient sleeps naturally. Patient instructions
include the following:
• Maintain regular
sleep-wake rhythm.
• Alcohol and sleeping
pills may alter the PSG results, but if they are
part of the patient's normal routine, they
should not be abruptly stopped.
• Avoid
stimulants, including medications for
narcolepsy.
• Avoid strenuous exercise on the
day of the PSG.
• Avoid naps on the day of
the sleep study.
Daytime PSG can be useful
for patients who typically sleep during the day.
Simplified sleep studies with limited subsets of
monitored parameters, such as PAP-NAPs, can be
used to help the patient with acclimatization
and finding optimal settings.
High costs and
long waiting lists have prompted the exploration
of alternative methods of evaluation and many
insurance companies are requiring home-based,
limited-channel sleep studies prior to
in-laboratory PSG. Instead of in-laboratory
titration, many patients with obstructive sleep
apnea can be started on automatically adjusting
continuous positive airway pressure (CPAP) and
then have the settings adjusted and response
monitored through data collected by the device.
Medicare guidelines
In 2008, Medicare
approved the use of unattended home sleep
monitoring devices of types II, III, or IV (with
at least 3 channels) if the patient received a
complete clinical evaluation and does not have
atypical or complicated symptoms and the studies
were read by a trained sleep specialist. The
guidelines for using a portable monitor
unattended home sleep study device for
continuous positive airway pressure (CPAP)
therapy include the following:
• Type II
device: This type of device has a minimum of 7
channels (eg, EEG, EOG, EMG, ECG-heart rate,
airflow, respiratory effort, oxygen saturation).
This type of device monitors sleep staging so
the apnea-plus-hypopnea index (AHI) can be
calculated.
• Type III device: This device
has a minimum of 4 channels, including
ventilation or airflow (at least 2 channels of
respiratory movement or airflow), heart rate or
ECG, and oxygen saturation.
• Type IV device:
This type of device does not meet requirements
for other types, and many measure only 1-2
parameters (eg, oxygen saturation or airflow).
For Medicare reimbursement, these devices,
including WatchPAT (Itamar Medical Ltd,
Caesarea, Israel) can be used if they have a
minimum of 3 channels.
American Academy of Sleep Medicine guidelines
The American Academy of Sleep Medicine (AASM)
evaluated the literature on unattended sleep
monitoring devices to develop their clinical
guidelines, published in 2007.[4] These
guidelines include the following recommendations
and cautions:
• Portable monitoring (PM) may
be indicated for the diagnosis of obstructive
sleep apnea (OSA) in patients for whom
in-laboratory PSG is not possible by virtue of
immobility, safety, or critical illness. PM may
also be indicated to monitor the non-CPAP
treatments of sleep apnea including oral
appliances, weight loss, and upper airway
surgery.
• PM is not appropriate for
diagnostic evaluation of patients who may have
comorbid sleep disorders including central sleep
apnea, periodic limb movements, insomnia,
parasomnias, circadian rhythm disorders, or
narcolepsy.
• PM is not appropriate for the
diagnosis of OSA in patients with significant
comorbid medical conditions that may degrade the
accuracy of PM. This includes, but is not
limited to, severe pulmonary disease, [5]
neuromuscular disease, or congestive heart
failure. PM is not indicated in the absence of a
comprehensive sleep evaluation.
• PM is not
appropriate for general screening of
asymptomatic patients.
• At minimum, PM must
record airflow, respiratory effort, and blood
oxygenation.
• The PM device must allow for
display of raw data with the capability of
manual scoring or editing of automated scoring
by a qualified sleep technologist.
• A
board-certified sleep specialist or an
individual who fulfills eligibility criteria for
the sleep medicine certification examination
must review the raw data from PM using scoring
criteria consistent with current published AASM
standards.
• False negative rates may be as
high as 17% in unattended PM studies. If the PM
test is technically inadequate or does not
provide the expected result, in-laboratory
polysomnography should be performed.
• AASM
does not support type IV devices for home sleep
testing.
In 2012, the AASM also published
evidence-based practice parameters for the
non-respiratory indications for polysomnography
and multiple sleep latency testing for
children.[6] PSG is indicated for children
suspected of having periodic limb movement
disorder (PLMD) for diagnosing PLMD. Children
with frequent NREM parasomnias, epilepsy, or
nocturnal enuresis should be clinically screened
for the presence of comorbid sleep disorders and
polysomnography should be performed if there is
a suspicion for sleep-disordered breathing or
PLMD.
Because of the lack of EEG monitoring,
Type III devices may underestimate the severity
of sleep-disordered breathing. Typically, events
must be associated with 3% desaturations to be
scored, so patients with events primarily
causing arousals may be missed. Additionally,
the apnea/hypopnea index (AHI) is calculated by
the number of apneas and hypopneas per hours of
test rather than hours of sleep, which can also
underestimate severity. For these reasons, if a
home study is normal in a patient with suspected
sleep apnea, an in-laboratory PSG is
recommended.
Multiple sleep latency test
Multiple sleep
latency testing (MSLT) is used to assess the
degree of daytime sleepiness and to evaluate for
possible narcolepsy. MSLT should be performed
after a full-night polysomnogram to ensure that
at least 6 hours of sleep precede the test and
that no other causes for excessive daytime
sleepiness are present. A sleep log should be
kept for at least 1 week prior to the study, and
all medications taken for the 2 weeks prior to
the study should be noted. Urine drug testing is
often done to evaluate for drugs that may affect
study results. Stimulant medications, nicotine,
and caffeine can affect the mean sleep latency,
and medications (especially selective serotonin
reuptake inhibitors [SSRIs]) can affect
sleep-onset rapid eye movement (REM) periods. In
general, SSRIs and stimulants need to be
discontinued at least 2 weeks prior to the test.
Small amounts of caffeine do not usually need to
be discontinued.
The patient is given 20
minute opportunities to nap every 2 hours for 4
or 5 naps. The first nap should begin within
1.5-3 hours after waking up. If the patient
falls asleep, he or she is allowed to sleep for
15 minutes. Sleep latency is the time to the
first epoch with over 15 seconds of any stage of
sleep. The mean sleep latency is determined. A
mean sleep latency of 10-15 minutes is
consistent with mild sleepiness, 5-10 minutes is
consistent with moderate sleepiness, and less
than 5 minutes is consistent with severe
sleepiness. A series of 2 sleep-onset REM
periods (SOREMP) is consistent with the
diagnosis of narcolepsy; however, only 80% of
patients with narcolepsy and 6.6% of patients
without narcolepsy had 2 or more SOREMPs in a
review of over 2000 MSLTs.
Maintenance of wakefulness test
Maintenance of wakefulness testing (MWT) is used
to determine how long a patient is able to
maintain wakefulness. The patient should be in a
dim room in a semirecumbent position. The 20-min
MWT includes 5 20-min tests of wakefulness, and
the 40-min MWT includes 4 40-min tests of
wakefulness every 2 hours. The first nap should
begin within 1.5-3 hours after waking up. A
preceding PSG is not always necessary. The
patient should be instructed to sit still and
remain awake as long as possible. The test ends
after 20 or 40 minutes if no sleep occurs or if
the patient achieves unequivocal sleep, which is
determined by either 3 consecutive epochs of
stage 1 sleep or 1 epoch of any other stage of
sleep. Sleep latency is the time to the first
epoch with over 15 seconds of any stage of
sleep. The mean sleep latency is determined.
Respiratory Events and Leg Movement Scoring
Basic rules
In 2015, the AASM updated scoring
rules, making changes to scoring of apneas,
hypopneas, Cheyne Stokes respiration, and
hypoventilation.
The event duration starts at
the nadir preceding the first breath that is
clearly reduced to the beginning of the first
breath that approximates the baseline breathing
amplitude. Events terminate if there is a clear
and sustained increase in breathing or if there
is a desaturation, when there is a re-saturation
of at least 2%.
Polysomnography reports
should report an apnea/hypopnea index (AHI),
which, for an in-lab study, is the number of
apneas and hypopneas per hour of sleep. For
portable studies, the AHI is the number of
apneas and hypopneas per hour of test. It is
important to know what criteria was used for the
events as there have been many changes to the
scoring criteria for hypopneas, which may have
led patients who were scored with desaturation
criteria only to have significantly
underestimated sleep apnea.
In general, an
AHI > 5 is considered significant sleep apnea.
Some polysomnography report respiratory
disturbance index (RDI), which is typically the
number of apneas plus hypopneas plus
respiratory-related arousals.
Obstructive apnea
See the list below:
•
Greater than 90% reduction in oronasal
thermistor amplitude for more than 10 seconds
• The duration of the 90% drop is greater than
or equal to 10 seconds
• Often associated
with increasing respiratory effort; usually seen
as paradoxical
Hypopnea
AASM recommendation
•
Reduction in nasal pressure amplitude by greater
than 30% lasting for at least 10 seconds
•
Associated with Sa02 drop of at least 3% from
pre-event baseline or an arousal
•
Obstructive hypopneas can be scored if any of
the following are present: Snoring, increased
inspiratory flattening of the nasal pressure, or
associated thoracoabdominal paradox during the
event but not pre-event
• Central hypopneas
can be scored if none of the above findings is
present Medicare definition
• Reduction in
pressure amplitude greater than 30% of the
baseline value
• Associated with Sa02
decrease of greater than 4%
Mixed apnea
See the list below:
•
Greater than 90% reduction in thermistor flow
for greater than 10 seconds
• Total absence
of respiratory effort at the beginning of the
event, followed by a gradual increase in effort,
which eventually breaks the apnea (usually
paradoxical)
Central apnea
See the list below:
•
Greater than 90% reduction in thermistor flow
for greater than 10 seconds
• Complete
absence of respiratory effort
Respiratory effort-related arousal
See the
list below:
• Greater than 10 breaths with
increasing respiratory effort or flattening of
the nasal pressure followed by an arousal
•
Does not meet criteria for hypopnea or apnea
Cheyne-Stokes respiration
See the list
below:
• At least 3 consecutive cycles
•
Cyclical crescendo and decrescendo change in
breathing amplitude
• Either 5 per hour of
sleep or duration greater than 10 minutes
•
Cycle length > 40 seconds
Sleep-related hypoventilation
See the list
below:
• Increase in C02 levels (ETC02 or
TCC02) to > 55 mmHg for greater than or equal to
10 minutes or
• Increase in C02 levels (ETC02
or TCC02) by at least 10 mmHg (from awake supine
value) to > 50 mmHg for greater than or equal to
10 minutes
Periodic limb movement
See the list below:
• Each jerk more than 0.5 seconds but less than
10 seconds in duration
• Minimum amplitude is
an 8 mV increase in EMG voltage above resting
EMG
• Must have 4 jerks separated by no less
than 5 seconds and no more than 90 seconds
•
No associated respiratory event within 0.5
seconds
EEG arousal
See the list below:
•
Abrupt shift of EEG frequency including alpha,
theta, and/or frequencies greater than 16 Hz
(but not spindles)
• Greater than 3 seconds
of changed frequency on EEG
• At least 10
seconds of stable sleep preceding the change
• In REM sleep, increase in submental EMG for at
least 1 second
Bruxism
See the list below:
• At least
twice the amplitude of baseline chin EMG
activity
• At least 3 elevations of 0.25-2
seconds of increased chin EMG activity
• One
elevation of greater than 2 seconds of increased
chin EMG activity
Disorders Evaluated With Polysomnography
Several different types of sleep disorders can
be evaluated and diagnosed using polysomnography
or sleep studies.
Dyssomnias (disorders of
initiating or maintaining sleep)
See the list
below:
• Circadian rhythm disorders
•
Narcolepsy
• Idiopathic hypersomnia
•
Inadequate sleep hygiene
• Sleep-wake
misperception
• Sleep-related respiratory
disorders
° Sleep apnea syndrome ° Upper
airway resistance syndrome ° Obesity
hypoventilation syndrome ° Central sleep apnea
Parasomnias
See the list below:
•
Disorders of arousal
• Disorders of
sleep-wake transition
• Disorders that occur
during REM sleep
o Nightmares
° REM
behavior disorder
• Medical-psychiatric sleep
disorders
° Medical - Sleep-related asthma
Please see also our Toxilact data base which is in the following language versions:
Toxilact Deutsche Sprachversion
Toxilact Nederlandstalige versie
Toxilakt έκδοση στην ελληνική γλώσσα
Toxilact English language version
Toxilact magyar nyelvű változat
Toxilact versione in lingua italiana
Toxilact polska wersja językowa
Pulmonary symptoms, findings and investigations
Assessment of chronic cough
Expectoration
Hemoptysis
Pleural Effusion
Approach to wheezing in children
Polysomnography
Pulmonary Function Testing
Toxicological risk during lactation
Toxicological lactation category I - the drug and/or its metabolites are either not eliminated through breast milk or are not toxic to the newborn and cannot lead to the development of absolutely any toxic reactions and adverse consequences for his health in the near and long term. Breast-feeding does not need to be discontinued while taking a given drug that falls into this toxicological lactation category.
Toxicological lactation category II - the drug and its metabolites are also eliminated through breast milk, but the plasma:milk ratio is very low and/or the excreted amounts cannot generate toxic reactions in the newborn due to various reasons, including degradation of the drug in the acid pool of the stomach of the newborn. Breastfeeding does not need to be discontinued while taking this medicine.
Toxicological lactation category III - the drug and/or its metabolites generate in breast milk equal to plasma concentrations or higher, and therefore the possible development of toxic reactions in the newborn can be expected. Breastfeeding should be discontinued for the period corresponding to the complete elimination of the drug or its metabolites from the mother's plasma.
Toxicological lactation category IV - the drug and/or its metabolites generate a plasma:milk ratio of 1:1 or higher and/or have a highly toxic profile for both the mother and the newborn, therefore their administration is incompatible with breastfeeding and it should to stop completely, and not just for the period of taking the drug, or to look for a less toxic therapeutic alternative.