Content

The content of medical knowledge in this section of the site of the Lactology Foundation is intended for the practical needs of doctors, pharmacists and students in these specialties. It is more than reasonable to consult other authoritative medical sources before using our medical knowledge.

Polysomnography

Nocturnal, laboratory-based polysomnography (PSG), also known as a sleep study, is the most commonly used test in the diagnosis of obstructive sleep apnea syndrome (OSAS).

Parameters Monitored
Assessment of sleep stages requires 3 studies: electroencephalography (EEG), electrooculography (EOG), and surface electromyography (EMG).
One EEG channel (central channel with an ear reference provides the best amplitude) is used to monitor sleep stage. However, most laboratories use 2 central channels and 2 occipital channels, with ear references as an adjunct to help identify sleep latency and arousals. A 10- to 20-electrode placement system is used to determine the location of these channels. Additional EEG channels can be used, particularly in patients with epilepsy (ie, a full 10-20 montage).
Two EOG channels are used to monitor both horizontal and vertical eye movements. Electrodes are placed at the right and left outer canthi, one above and one below the horizontal eye axis. The electrodes pick up the inherent voltage within the eye; the cornea has a positive charge and the retina has a negative charge. Evaluation of the eye movements is necessary for 2 reasons. First is for documentation of the onset of rapid eye movement (REM) sleep, and second is to note the presence of slow-rolling eye movements that usually accompany the onset of sleep.
One EMG channel (usually chin or mentalis and/or submentalis) is used to record atonia during REM sleep or lack of atonia in patients with REM-related parasomnias. To assess bruxism, the EMG electrodes can be placed over the masseter. The EMG recording from other muscle groups is assessed for other sleep disorders. For example, the anterior tibialis EMG is helpful for assessing periodic limb movements during sleep and the intercostal EMG is used as adjunctive help for determining effort during respiratory events.
Two channels are used for monitoring airflow. One thermistor channel (oral and/or nasal) is used to evaluate the presence or absence of airflow. Any change in temperature as a patient inhales and exhales leads to a normal signal, so this channel is insensitive to partial flow obstruction. Thermistor is the recommended channel for evaluation of apneas. Nasal pressure transducer channel is a more sensitive measure of airflow restriction. Normal breathing has a rounded pattern, while resistance to airflow leads to a squaring off of the flow signal. Pressure transducer is the recommended channel for evaluating hypopneas. It is also used for airflow resistance in upper airway resistance syndrome. Other parameters that can be monitored in a sleep study include the following:
• Electrocardiography
• Pulse oximetry
• Respiratory effort (thoracic and abdominal)
• End tidal or transcutaneous C02
• Sound recordings to measure snoring
• Surface EMG monitoring of limb muscles (to detect limb movements, periodic or other)
• Continuous video monitoring
Optional parameters that can be monitored in a sleep study include the following:
• Core body temperature
• Incident light intensity
• Penile tumescence
• Pressure and pH at various esophageal levels

Staging of Sleep
Standardized criteria for the staging of sleep were published first in 1968 by Rechtschaffen and Kales. A revised version was published in 2007 by the American Academy of Sleep Medicine. The chief revision was the consolidation of stages 3 and 4 into a single stage N3 (slow wave sleep). Previous stages 1 and 2 were renamed N1 and N2. Both systems are reflected below, with the expectation that the discontinued stages will be omitted from future revisions of this article.

EEG background
Alpha EEG
• Frequency of 8-13 Hz
• Produced in the occipital region
• Crescendo-decrescendo appearance Theta EEG
• Frequency of 3-7 Hz
• Produced in the central vertex region
• No amplitude criteria
• Most common sleep frequency Delta EEG
• EEG frequency of 0.5-2 Hz
• Seen predominantly in the frontal region
• Amplitude of greater than 75 microvolts

Theta EEG
• Frequency of 3-7 Hz
• Produced in the central vertex region
• No amplitude criteria
• Most common sleep frequency Delta EEG
• EEG frequency of 0.5-2 Hz
• Seen predominantly in the frontal region
• Amplitude of greater than 75 microvolts

Delta EEG
• EEG frequency of 0.5-2 Hz
• Seen predominantly in the frontal region
• Amplitude of greater than 75 microvolts

Sleep spindle
See the list below:
• Frequency of 12-14 Hz
• Produced in central-vertex region
• Greater than 0.5-3 seconds in duration
• 0.5-second spindles with 6-7 cycles
• Indicative of stage 2 sleep

K complexes
See the list below:
• Sharp, slow waves with a negative, then positive, deflection
• No amplitude criteria
• Duration must be at least 0.5 seconds
• Predominantly produced in central-vertex region
• Indicative of stage 2 sleep
• May occur with or without stimuli

Wake stage
See the list below:
• Greater than 50% of each epoch contains alpha activity
• Eye blinks at a frequency of 0.5-2 Hz
• Reading eye movements
• Irregular conjugate rapid eye movements associated with normal or high chin tone.

Stage N1 (formerly stage 1)
See the list below:
• Greater than 50% of the epoch contains theta activity (4-7 Hz) with slowing of the background rhythms greater than or equal to 1 Hz from those of stage wake
• Vertex sharp waves
• Slow-rolling eye movements in EOG channels
• Relatively high submental EMG tone

Stage N2 (formerly stage 2)
See the list below:
• Theta activity (4-7 Hz)
• K-complexes and sleep spindles occur episodically
• High tonic submental EMG

Stage N3
See the list below:
• Greater than 20% of each epoch must contain delta activity
• Amplitude of 75 microvolts or greater
• Submental muscle tone may be slightly reduced

Discontinued former stage 3
See the list below:
• Between 20-50% of each epoch must contain delta activity
• Amplitude of 75 microvolts or greater
• Submental muscle tone may be slightly reduced

Discontinued former stage 4
See the list below:
• Greater than 50% of the epoch has scorable delta activity
• Amplitude of 75 microvolts or greater
• Submental EMG activity slightly reduced from that of light sleep

REM sleep
See the list below:
• Rapid eye movements
• Low amplitude, mixed frequency EEG (similar to awake pattern)
• Atonia or the lowest tonic submental EMG
• May see saw-tooth waves

Procedures
In 1992, the Office of Technology Assessment of the Agency of Health Care Policy and Research recommended, in an evidence-based assessment, declared two tests as having been studied sufficiently. Both tests are performed in a sleep laboratory.
The first is overnight polysomnography (PSG) or sleep study, which is an overnight recording of the patient's sleep. Typically for a baseline study, the patient is observed sleeping naturally without any treatment, but if a significant amount of sleep-disordered breathing (AHI> 20-30 events per hour) is seen in the first hours of the study, a split-night study is performed during which positive airway pressure (PAP) is started. Titration studies may also be done with initiation of PAP from the start of the study to determine optimal settings.
The second is multiple sleep latency testing (MSLT), which records multiple naps throughout a day. Maintenance of wakefulness testing (MWT) can also be performed, which determines how long wakefulness can be maintained.
Standard sleep studies usually use the overnight PSG (may be performed over several nights). If daytime sleepiness is an issue and cannot be fully explained by the overnight study results, an MSLT should be performed the next day. Limitations usually stem from the fact that recording conditions may not reflect what happens during a regular night in the patient's home.
Although diagnosing a sleep problem on the basis of a recording over a single night is common practice, some authorities caution that more than one night of recording may be necessary so the patient can become comfortable with unfamiliar surroundings and sleep more naturally. This effect is greatest on the first night in the sleep laboratory (ie, first-night effect).
Sporadic events may be missed with a single-night PSG. External factors that disturb the subject's sleep may be present in the home but absent from the controlled environment of the sleep laboratory.

Patient preparation is important so that the patient sleeps naturally. Patient instructions include the following:
• Maintain regular sleep-wake rhythm.
• Alcohol and sleeping pills may alter the PSG results, but if they are part of the patient's normal routine, they should not be abruptly stopped.
• Avoid stimulants, including medications for narcolepsy.
• Avoid strenuous exercise on the day of the PSG.
• Avoid naps on the day of the sleep study.
Daytime PSG can be useful for patients who typically sleep during the day. Simplified sleep studies with limited subsets of monitored parameters, such as PAP-NAPs, can be used to help the patient with acclimatization and finding optimal settings.
High costs and long waiting lists have prompted the exploration of alternative methods of evaluation and many insurance companies are requiring home-based, limited-channel sleep studies prior to in-laboratory PSG. Instead of in-laboratory titration, many patients with obstructive sleep apnea can be started on automatically adjusting continuous positive airway pressure (CPAP) and then have the settings adjusted and response monitored through data collected by the device.

Medicare guidelines
In 2008, Medicare approved the use of unattended home sleep monitoring devices of types II, III, or IV (with at least 3 channels) if the patient received a complete clinical evaluation and does not have atypical or complicated symptoms and the studies were read by a trained sleep specialist. The guidelines for using a portable monitor unattended home sleep study device for continuous positive airway pressure (CPAP) therapy include the following:
• Type II device: This type of device has a minimum of 7 channels (eg, EEG, EOG, EMG, ECG-heart rate, airflow, respiratory effort, oxygen saturation). This type of device monitors sleep staging so the apnea-plus-hypopnea index (AHI) can be calculated.
• Type III device: This device has a minimum of 4 channels, including ventilation or airflow (at least 2 channels of respiratory movement or airflow), heart rate or ECG, and oxygen saturation.
• Type IV device: This type of device does not meet requirements for other types, and many measure only 1-2 parameters (eg, oxygen saturation or airflow). For Medicare reimbursement, these devices, including WatchPAT (Itamar Medical Ltd, Caesarea, Israel) can be used if they have a minimum of 3 channels.

American Academy of Sleep Medicine guidelines
The American Academy of Sleep Medicine (AASM) evaluated the literature on unattended sleep monitoring devices to develop their clinical guidelines, published in 2007.[4] These guidelines include the following recommendations and cautions:
• Portable monitoring (PM) may be indicated for the diagnosis of obstructive sleep apnea (OSA) in patients for whom in-laboratory PSG is not possible by virtue of immobility, safety, or critical illness. PM may also be indicated to monitor the non-CPAP treatments of sleep apnea including oral appliances, weight loss, and upper airway surgery.
• PM is not appropriate for diagnostic evaluation of patients who may have comorbid sleep disorders including central sleep apnea, periodic limb movements, insomnia, parasomnias, circadian rhythm disorders, or narcolepsy.
• PM is not appropriate for the diagnosis of OSA in patients with significant comorbid medical conditions that may degrade the accuracy of PM. This includes, but is not limited to, severe pulmonary disease, [5] neuromuscular disease, or congestive heart failure. PM is not indicated in the absence of a comprehensive sleep evaluation.
• PM is not appropriate for general screening of asymptomatic patients.
• At minimum, PM must record airflow, respiratory effort, and blood oxygenation.
• The PM device must allow for display of raw data with the capability of manual scoring or editing of automated scoring by a qualified sleep technologist.
• A board-certified sleep specialist or an individual who fulfills eligibility criteria for the sleep medicine certification examination must review the raw data from PM using scoring criteria consistent with current published AASM standards.
• False negative rates may be as high as 17% in unattended PM studies. If the PM test is technically inadequate or does not provide the expected result, in-laboratory polysomnography should be performed.
• AASM does not support type IV devices for home sleep testing.
In 2012, the AASM also published evidence-based practice parameters for the non-respiratory indications for polysomnography and multiple sleep latency testing for children.[6] PSG is indicated for children suspected of having periodic limb movement disorder (PLMD) for diagnosing PLMD. Children with frequent NREM parasomnias, epilepsy, or nocturnal enuresis should be clinically screened for the presence of comorbid sleep disorders and polysomnography should be performed if there is a suspicion for sleep-disordered breathing or PLMD.
Because of the lack of EEG monitoring, Type III devices may underestimate the severity of sleep-disordered breathing. Typically, events must be associated with 3% desaturations to be scored, so patients with events primarily causing arousals may be missed. Additionally, the apnea/hypopnea index (AHI) is calculated by the number of apneas and hypopneas per hours of test rather than hours of sleep, which can also underestimate severity. For these reasons, if a home study is normal in a patient with suspected sleep apnea, an in-laboratory PSG is recommended.

Multiple sleep latency test
Multiple sleep latency testing (MSLT) is used to assess the degree of daytime sleepiness and to evaluate for possible narcolepsy. MSLT should be performed after a full-night polysomnogram to ensure that at least 6 hours of sleep precede the test and that no other causes for excessive daytime sleepiness are present. A sleep log should be kept for at least 1 week prior to the study, and all medications taken for the 2 weeks prior to the study should be noted. Urine drug testing is often done to evaluate for drugs that may affect study results. Stimulant medications, nicotine, and caffeine can affect the mean sleep latency, and medications (especially selective serotonin reuptake inhibitors [SSRIs]) can affect sleep-onset rapid eye movement (REM) periods. In general, SSRIs and stimulants need to be discontinued at least 2 weeks prior to the test. Small amounts of caffeine do not usually need to be discontinued.
The patient is given 20 minute opportunities to nap every 2 hours for 4 or 5 naps. The first nap should begin within 1.5-3 hours after waking up. If the patient falls asleep, he or she is allowed to sleep for 15 minutes. Sleep latency is the time to the first epoch with over 15 seconds of any stage of sleep. The mean sleep latency is determined. A mean sleep latency of 10-15 minutes is consistent with mild sleepiness, 5-10 minutes is consistent with moderate sleepiness, and less than 5 minutes is consistent with severe sleepiness. A series of 2 sleep-onset REM periods (SOREMP) is consistent with the diagnosis of narcolepsy; however, only 80% of patients with narcolepsy and 6.6% of patients without narcolepsy had 2 or more SOREMPs in a review of over 2000 MSLTs.

Maintenance of wakefulness test
Maintenance of wakefulness testing (MWT) is used to determine how long a patient is able to maintain wakefulness. The patient should be in a dim room in a semirecumbent position. The 20-min MWT includes 5 20-min tests of wakefulness, and the 40-min MWT includes 4 40-min tests of wakefulness every 2 hours. The first nap should begin within 1.5-3 hours after waking up. A preceding PSG is not always necessary. The patient should be instructed to sit still and remain awake as long as possible. The test ends after 20 or 40 minutes if no sleep occurs or if the patient achieves unequivocal sleep, which is determined by either 3 consecutive epochs of stage 1 sleep or 1 epoch of any other stage of sleep. Sleep latency is the time to the first epoch with over 15 seconds of any stage of sleep. The mean sleep latency is determined.

Respiratory Events and Leg Movement Scoring
Basic rules
In 2015, the AASM updated scoring rules, making changes to scoring of apneas, hypopneas, Cheyne Stokes respiration, and hypoventilation.
The event duration starts at the nadir preceding the first breath that is clearly reduced to the beginning of the first breath that approximates the baseline breathing amplitude. Events terminate if there is a clear and sustained increase in breathing or if there is a desaturation, when there is a re-saturation of at least 2%.
Polysomnography reports should report an apnea/hypopnea index (AHI), which, for an in-lab study, is the number of apneas and hypopneas per hour of sleep. For portable studies, the AHI is the number of apneas and hypopneas per hour of test. It is important to know what criteria was used for the events as there have been many changes to the scoring criteria for hypopneas, which may have led patients who were scored with desaturation criteria only to have significantly underestimated sleep apnea.
In general, an AHI > 5 is considered significant sleep apnea. Some polysomnography report respiratory disturbance index (RDI), which is typically the number of apneas plus hypopneas plus respiratory-related arousals.

Obstructive apnea
See the list below:
• Greater than 90% reduction in oronasal thermistor amplitude for more than 10 seconds
• The duration of the 90% drop is greater than or equal to 10 seconds
• Often associated with increasing respiratory effort; usually seen as paradoxical

Hypopnea
AASM recommendation
• Reduction in nasal pressure amplitude by greater than 30% lasting for at least 10 seconds
• Associated with Sa02 drop of at least 3% from pre-event baseline or an arousal
• Obstructive hypopneas can be scored if any of the following are present: Snoring, increased inspiratory flattening of the nasal pressure, or associated thoracoabdominal paradox during the event but not pre-event
• Central hypopneas can be scored if none of the above findings is present Medicare definition
• Reduction in pressure amplitude greater than 30% of the baseline value
• Associated with Sa02 decrease of greater than 4%

Mixed apnea
See the list below:
• Greater than 90% reduction in thermistor flow for greater than 10 seconds
• Total absence of respiratory effort at the beginning of the event, followed by a gradual increase in effort, which eventually breaks the apnea (usually paradoxical)

Central apnea
See the list below:
• Greater than 90% reduction in thermistor flow for greater than 10 seconds
• Complete absence of respiratory effort

Respiratory effort-related arousal
See the list below:
• Greater than 10 breaths with increasing respiratory effort or flattening of the nasal pressure followed by an arousal
• Does not meet criteria for hypopnea or apnea

Cheyne-Stokes respiration
See the list below:
• At least 3 consecutive cycles
• Cyclical crescendo and decrescendo change in breathing amplitude
• Either 5 per hour of sleep or duration greater than 10 minutes
• Cycle length > 40 seconds

Sleep-related hypoventilation
See the list below:
• Increase in C02 levels (ETC02 or TCC02) to > 55 mmHg for greater than or equal to 10 minutes or
• Increase in C02 levels (ETC02 or TCC02) by at least 10 mmHg (from awake supine value) to > 50 mmHg for greater than or equal to 10 minutes

Periodic limb movement
See the list below:
• Each jerk more than 0.5 seconds but less than 10 seconds in duration
• Minimum amplitude is an 8 mV increase in EMG voltage above resting EMG
• Must have 4 jerks separated by no less than 5 seconds and no more than 90 seconds
• No associated respiratory event within 0.5 seconds

EEG arousal
See the list below:
• Abrupt shift of EEG frequency including alpha, theta, and/or frequencies greater than 16 Hz (but not spindles)
• Greater than 3 seconds of changed frequency on EEG
• At least 10 seconds of stable sleep preceding the change
• In REM sleep, increase in submental EMG for at least 1 second

Bruxism
See the list below:
• At least twice the amplitude of baseline chin EMG activity
• At least 3 elevations of 0.25-2 seconds of increased chin EMG activity
• One elevation of greater than 2 seconds of increased chin EMG activity

Disorders Evaluated With Polysomnography
Several different types of sleep disorders can be evaluated and diagnosed using polysomnography or sleep studies.
Dyssomnias (disorders of initiating or maintaining sleep)
See the list below:
• Circadian rhythm disorders
• Narcolepsy
• Idiopathic hypersomnia
• Inadequate sleep hygiene
• Sleep-wake misperception
• Sleep-related respiratory disorders
° Sleep apnea syndrome ° Upper airway resistance syndrome ° Obesity hypoventilation syndrome ° Central sleep apnea
Parasomnias
See the list below:
• Disorders of arousal
• Disorders of sleep-wake transition
• Disorders that occur during REM sleep
o Nightmares
° REM behavior disorder
• Medical-psychiatric sleep disorders
° Medical - Sleep-related asthma

Please see also our Toxilact data base which is in the following language versions:

نسخة اللغة العربية Toxilact

Toxilact中文版

Toxilact česká jazyková verze

Toxilact dansk sprogversion

Toxilact Deutsche Sprachversion

Toxilact Nederlandstalige versie

Toxilakt έκδοση στην ελληνική γλώσσα

Toxilact English language version

Version française de Toxilact

Toxilact magyar nyelvű változat

Toxilact versione in lingua italiana

トキシラクト 日本語版

גרסת השפה הישראלית רעילה

Toxilact norsk språkversjon

Toxilact polska wersja językowa

Toxilact versão em português

Токсилак русскоязычная версия

Toxilac versión en idioma español

Toxilact svensk språkversion

Toxilact Türkçe dil versiyonu

If our cause of developing a less toxic world and healthier babies in it appeals to you, you can support us with a donation!

Pulmonary symptoms, findings and investigations

Assessment of chronic cough

Expectoration

Hemoptysis

Pleural Effusion

Approach to wheezing in children

Polysomnography

Pulmonary Function Testing


Toxicological risk during lactation

Toxicological lactation category I - the drug and/or its metabolites are either not eliminated through breast milk or are not toxic to the newborn and cannot lead to the development of absolutely any toxic reactions and adverse consequences for his health in the near and long term. Breast-feeding does not need to be discontinued while taking a given drug that falls into this toxicological lactation category.

Toxicological lactation category II - the drug and its metabolites are also eliminated through breast milk, but the plasma:milk ratio is very low and/or the excreted amounts cannot generate toxic reactions in the newborn due to various reasons, including degradation of the drug in the acid pool of the stomach of the newborn. Breastfeeding does not need to be discontinued while taking this medicine.

Toxicological lactation category III - the drug and/or its metabolites generate in breast milk equal to plasma concentrations or higher, and therefore the possible development of toxic reactions in the newborn can be expected. Breastfeeding should be discontinued for the period corresponding to the complete elimination of the drug or its metabolites from the mother's plasma.

Toxicological lactation category IV - the drug and/or its metabolites generate a plasma:milk ratio of 1:1 or higher and/or have a highly toxic profile for both the mother and the newborn, therefore their administration is incompatible with breastfeeding and it should to stop completely, and not just for the period of taking the drug, or to look for a less toxic therapeutic alternative.